Key words
|
| UV- Visible Spectrophotometer, Curcumin, ICHguidelines |
INTRODUCTION
|
| The powdered dry rhizome of the plant Curcumalonga, commonly called turmeric, is widely used as acolouring agent and spice in many food items [1]. Itcontains a wide variety of phytochemicals, including curcumin, demethoxycurcumin,bisdemethoxycurcumin, zingiberene, curcumenol,curcumol, eugenol, tetrahydrocurcumin,triethylcurcumin, turmerin, turmerones, andturmeronols [2, 3]. Curcumin is the phytochemical thatgives a yellow color to turmeric and is nowrecognized as being responsible for most of thetherapeutic effects [4]. Chemically described as (1E,6E)-1.7-bis (4 hydroxy – 3 methoxyphenyl) – 1,6Heptadiene-3,5-dione, the aromatic ring systems,which are polyphenols are connected by two α, β –unsaturated carbonyl groups (Fig. 1), while the α,β –unsaturated carbonyl is a good Michael acceptor andundergoes nucleophilic addition [5, 6]. It ishydrophobic in nature and frequently soluble indimethylsulfoxide, acetone, ethanol, and oils. It hasan absorption maximum around 420 nm [7, 8]. |
| Literature survey revealed that a variety of analyticalmethods viz. HPLC, HPTLC, UV-Visible has beendeveloped for their analysis but in plasma and urine[9, 10]. As the formulations are available withoutcombinations of any drugs, there is a need for comingup with analytical method which is simple, sensitive,rapid and accurate for estimation of Curcumin inpure form and in pharmaceutical preparations [11, 12].Therefore, the aim of the present work is to developand validate a method for the analysis by UV-Visiblespectrophotometer which is easily adaptable as aroutine in quality testing laboratories. This hasenabled us to reduce total time of analysis besidestaking care of the error caused due to |
MATERIALS AND METHODS
|
Materials
|
| Curcumin was obtained from Loba chieme Mumbai,India as gift sample. Methanol used was of analyticalgrade and purchased form Merk Chemicals, India.Three formulations collected from market (A, B & C)with drug equivalent to 500 mg curcumin. All the other chemicals and reagents used were of analyticalgrade. |
Method development
|
Instrumentation
|
| Spectroscopic analysis was carried out using DoublebeamShimadzu recording UV-VisibleSpectrophotometer (Kyoto, Japan) model 1601 with10 mm path length matched quartz cells was used foranalytical purpose. |
Standard stock solution
|
| Stock solutions of curcumin containing 10μg/ml wereprepared in methanol and its aliquots weretransferred in a series of 10 ml volumetric flasks invarying fractions and their volumes were made withmethanol to prepare different standard dilutionsvarying in between 1-7mg ml-1. |
Method optimization
|
Selection and Optimization of Solvent
|
| It is well known that the solvent do exerts a profoundinfluence on the quality and shape of the peak [13].The choices of solvents for UV method developmentare: Chloroform, Acetone, Methanol etc. Different ofsolvents were optimized. Out of which methanolsatisfied all the conditions relative to Peak quality &non-interference at the specified wavelength. |
Selection of Wavelength
|
| The wavelength at which maximum absorption takesplace in UV detector is selected for further analysisi.e. 421nm. |
Method validation
|
| Validation of the method was carried out as per theInternational Conference on Harmonization (ICH)guidelines Q2 (R1) (ICH, 2005) [14]. And accordinglythe parameters evaluated were: |
Sensitivity
|
| Sensitivity of the method was determined withrespect to limit of detection (LOD) and limit of quantification (LOQ) [15]. Series of concentrations ofdrug solutions (0.01–7 μg/ml) were used andanalyzed to determine LOD and LOQ. |
| LOD and LOQ were experimentally verified bydiluting known concentration of Curcumin until theaverage responses were approximately 3 or 10 timesthe standard deviation of the responses for sixreplicate determinations [16]. |
Specificity and selectivity
|
| Three different marketed tablets of Curcumin ofconcentration 4 mg ml-1 were prepared in methanoland 4 mg ml-1 of standard Curcumin were analyzed bythe proposed method. The estimated amounts ofmarketed formulation were compared with that ofpure Curcumin solution of the same strength. |
Linearity and range
|
| Seven different concentrations (1-7mg ml-1) ofCurcumin were prepared in methanol from a freshstock of 10 mg ml-1.Least square regression analysiswas done for the obtained data. |
Accuracy
|
| In standard analysis method, three differentconcentrations of the standard Curcumin inmethanol were prepared (2.5 mg ml-1, 4 mg ml-1& 5.5mg ml-1) from independent stock solutions and theirstrengths were estimated by the standard curve.Standard addition method was followed to supportthe accuracy by adding separately three differentstandard concentrations of Curcumin (0.5 mg ml-1, 1mg ml-1 and 1.5 mg ml-1) to a pre-analyzed Curcuminsolution of 4 mg ml-1 and analyzing them again in thesame way. The accuracy was reported as % recovery ±(% confidence interval) with % relative error on thebase of actual and estimated concentrations. |
Precision
|
| Repeatability was done by analyzing three differentconcentrations of Curcumin( 2.5mg ml-1, 4 mg ml-1and 5.5 mg ml-1) in methanol in six let on a single day.Intermediate precision was done by analyzing thesame three concentrations on three different days in six let (drug was found stable for three days).Reproducibility was determined by analyzing threedifferent concentrations of Curcumin (2.5 mg ml-1, 4mg ml-1 and 5.5 mg ml-1) in six let on different UVspectrophotometers (One Cintra 5 double beam UVSpectrophotometer and two different Shimadzu 1601double beam Spectrophotometers in different labs).% Relative standard deviation, standard deviationand confidence interval of the estimatedconcentrations based on standard curve werereported for each set of data. |
Robustness
|
| Robustness of the proposed method was alsodetermined by changing the λ max of the analysis (λ max420 nm) by ± 1.0 nm. % Mean recovery (± %confidence interval) as well as % relative error wasreported. |
Use of above method for marketedformulations
|
| The content of Curcumin in tablets (labelled claim:500 mg per tablet) were determined by powderingtwenty tablets and powder equivalent to 10 mg ofCurcumin was weighed. The drug from the powderwas extracted with methanol (Hanif et al., 1997). Forcomplete extraction of the drug, it was sonicated for30 min and volume was made up to 100 ml. Theresulting solution was centrifuged at 2500 rpm for10 min and supernatant was analyzed for drugcontent. |
| Three different marketed tablets of Curcumin wereused to prepare three independent stocks ofCurcumin in methanol of 500 mg ml-1 concentration.These three stocks were used individually to preparethree different concentrations of Curcumin (2.5 mgml-1, 4 mg ml-1 and 5.5 mg ml-1). The preparedsolutions were assayed by the proposed method. The% assay values with % confidence intervals arereported. |
RESULTS AND DISCUSSION
|
| The proposed method was found to be simple,sensitive, accurate, precise, economical and rapid forthe routine estimation of Curcumin in bulk drug andpharmaceutical dosage forms. |
Analysis of the Drug
|
| Melting point of curcumin was found to be 183°C.Drug was freely soluble in methanol, chloroform,ethanol, acetone and practically insoluble in water.Spectral scan – λmax of curcumin was found to be at421 nm. |
Analytical Method ValidationParameters
|
| The method was validated as per ICH guidelines (Q2(R1). |
Linearity and Range
|
| Linearity of an analytical method is its ability, withina given range, to obtain test results that are directly,or through a mathematical transformation,proportional to concentration of analyte [17]. Goodlinear correlations were obtained betweenabsorbance and concentration in the selected rangeof 1 – 7 μg/ml. Characteristic parameters are Slope ±S.D. 0.1265 ± 0.15, Intercept ± S.D. 0.0174 ± 0.27,regression coefficient of 0.9991 and correlationcoefficient of 0.9995 between the standard drugconcentration and corresponding mean absorbanceshow a good linearity of standard curve (Table 1). |
Precision
|
| The precision of an analytical method expresses thedegree of scatter between a series of measurementsobtained from multiple sampling of the samehomogeneous sample under prescribed conditions[18]. Intraday precision refers to the use of analyticalprocedure within a laboratory over a short period oftime using the same operator with the sameequipment whereas Interday precision involvesestimation of variations in analysis when a method isused within a laboratory on different days, by different analysts [19, 20]. Repeatability (intraday) wasassessed by analyzing these three differentConcentrations (2.5, 4.0, 5.5 μg/ml), three times aday. Intermediate precision (Interday) wasestablished by analyzing these three differentconcentrations(2.5, 4.0, 5.5 μg/ml), three times a dayfor at least three different days (Table 2). |
| The Standard Deviation, % RSD and ConfidenceInterval for the intra-assay precision, intermediateprecision and reproducibility for all the threeconcentration levels were found below 0.018, 0.495,± 0.014 & 0.016, 0.570, ± 0.013 and 0.09, 100.006, ±0.072 respectively. The data indicated above showedan excellent intraday precision, intermediateprecision and reproducibility of the proposedmethod. |
Accuracy
|
| Accuracy of an analytical method is the closeness oftest results to true value. It was determined by theapplication of analytical procedure to recoverystudies, where known amount of standard is spikedin preanalyzed samples solutions [21, 22]. |
| The % recovery for the standard analysis andreference analysis method for all the threeconcentration levels ranged from 99.1% to 101.4%with confidence interval ranging from ± 0.090 to ±0.190 showing that any small change in the drugconcentration can be accurately determined withhigh accuracy. The results obtained form thestandard addition and reference analysis methodware also found supporting the accuracy of theproposed method. |
Specificity
|
| The presence of excipients in formulation does notinterfere with the drug peak. Therefore, the proposedmethod was found specific and selective for the drug. |
LOD/LOQ
|
| LOD and LOQ were calculated according to theformulae: |
| LOD=3.3 σ / S =0.05mg/ml |
| LOQ=10 σ / S = 0.1724mg/ml |
Robustness
|
| The variation in the λ max within limits ± 1.0 nmbrought % recovery lying in between 99.0 to 99.7with a maximum % confidence interval of ± 0.009,indicating it to be a sufficiently robust method. |
Application of the Validated UVVisibleSpectrophotometer Method on theMarketed Formulation
|
| The marketed tablet formulations were analyzed bythe proposed method. In accordance with ICHguidelines the assay values for all these formulationswere found to be ranging in between 99.89 to 100.19with a maximum % confidence interval of ± 0.11(Table 3). |
CONCLUSION
|
| The analytical method developed on UV- VisibleSpectrophotometer was simple, reliable, accurate andreproducible. The method eliminates extraction stepsthus reduce analytical time, cost and minimize theextraction errors. Low cost, faster speed, satisfactoryprecision and good specificity, to assessunequivocally the analyte in the presence ofcomponents, which may be expected to be present,are the main features of this method. Method wassuccessfully validated as per ICH guidelines and canbe conveniently employed for routine quality controlanalysis of Curcumin in bulk drug, marketed tabletsand other formulations without any interferencefrom excipients. The method was comparable to theexisting methods in all respects, which analyze thedrugs but in plasma. |
Conflict of Interest
|
| NIL |
Source of Support
|
| NONE |
ACKNOWLEDGEMENTS
|
| The authors thank Loba Chieme and Sami ChemicalsLtd, Mumbai for the gift samples of Curcumin. Theauthors are also grateful to Prof. (Dr.) Shyam S.Agrawal (Professor, Pharmacology, AMITY) and Dr.Monica Gupta (Lecturer, Pharmaceutical Chemistry, DIPSAR), respectively, for their valuable suggestions,active guidance and facilities provided duringresearch work. |
| |
Tables at a glance
|
|
|
| |
Figures at a glance
|
 |
| Figure 1 |
|
| |
References
|
- Hanif R, Qiao L, Shiff SJ, Rigas B 1997. Curcumin, a natural plant phenolic food additive, inhibits cell proliferation and induces cell cycle changes in colon adenocarcinoma cell lines by a prostaglandinindependent pathway. J. Lab. Clin. Med. 130: 576- 584.
- Almedia LP, Cherubino, APF, Alves RJ, Dufosse L, Gloria MBA 2005. Separation and determination ofthe physico-chemical characteristics of curcumin, demethoxycurcumin and bisdemethoxycurcumin. Food Res Int. 38: 1039-1044.
- Abas F, Lajis NH, Shaari K, Israf DA, Stanslas J, Yusuf UK, Raof SM 2005. A labdane diterpene glucoside from the rhizomes of Curcuma mangga. J Nat Prod 68: 1090–1093.
- Bharti AC, Donato N, Singh S, Aggarwal BB 2003. Curcumin (diferuloylmethane) down-regulates the constitutive activation of nuclear factor-Kappa B and IKappa-Balpha kinase in human multiple myeloma cells, leading to suppression of proliferation and induction of apoptosis. Blood 101: 1053-1062.
- Chan MM 1995. Inhibition of tumor necrosis factor by curcumin, a phytochemical. Biochem Pharmacol 49:1551-1556.
- Donatus I A, Sardjoko, Vermeulen N P 1990. Cytotoxic and cytoprotective activities of curcumin - Effects on paracetamol-induced cytotoxicity, lipid peroxidation and glutathione depletion in rat hepatocytes. Biochem. Pharmacol. 39: 1869-1875.
- Heath DP, Pruitt MA, Brenner DE, Rock CL 2003. Curcumin in plasma and urine: quantiation by high performance liquid chromatography. J Chrom B. 783: 287-95.
- Mandal V, Mohan Y, Hemalatha S 2007. Optimization of curcumin extraction by microwave assisted in vitro plant cell bursting by orthogonal array designed extraction process and HPTLC analysis. Phcog Mag. 3:132-138.
- Somasundaram S, Edmund NA, Moore DT, Small GW, Shi YY, Orlowski R Z 2002. Dietary curcumin inhibits chemotherapy-induced apoptosis in models of human breast cancer. Cancer Res. 62: 3868- 3875.
- Srinivasan KR 1953. A chromatographic study of the curcuminoids in curcuma longa Linn. J. Pharm. Pharmacol. 5: 448-453.
- Marsin SM, Ahmad UK, Smith RM 1993. Application of supercritical fluid extraction and chromatography to the analysis of turmeric. J Chromatogr Sc. 31: 20-25.
- Nagabhushan M, Bhide SV 1992. Curcumin as an inhibitor of cancer. J. Am. Coll. Nutr. 11: 192-198.
- Ahuja S, Scypinsk S. 2001. Handbook of modern pharmaceutical analysis. 5th ed., London: Academic Press. p 345-442.
- ICH Guideline Q2(R1), Validation of analytical procedures: text and methodology, November 2005.
- Bailey, L.C., 2001. Chromatography, in: Gennaro, A.R., (Eds.), Remington: The science and practice of pharmacy. 20th ed. Philadelphia: Lippincott Williams & Wilkins, p. 587-613.
- Carr GP 1990. A Parallel Approach to Method Validation in Pharmaceutical Analysis. J Pharm Biomed Anal 8: 613-618.
- Chapman KG 1993. Validation terminology. In Reddy IR, Nash RA, editors. Pharmaceutical process validation, 2nd ed., New York: Maecel Dekker. p 587-596.
- Chatwal GR, Anand SK. 2007. Instrumental methods of chemical analysis, 5th ed. Gurgaon: Himalaya Publishing House. p 2.613-2.615.
- Connors AK. 1999. A textbook of Pharmaceutical Analysis, 3rd ed. New York: Interscience Publications. p 581 - 585.
- Helmut G, Alex W. 2001. Hand Book of Analytical Techniques, 3rd ed., New York: Wiley InterScience. p 283-326.
- Lindsay S. 1987. HPLC Analytical chemistry by open learning, 4th ed., New York: John Wiley &Sons. p 54-72.
- Snyder LR, Kirkland JJ, Glajch JL. 1997. Completing the method: Validation and transfer. In Snyder LR, Kirkland JJ, Glajch JL, editors. Practical HPLC method development, 2nd ed., New York: John Wiley & Sons. p 233-312.
|